3152B User Manual Publication No. 980935 Rev. M
7-2 Adjustments and Firmware Update Astronics Test Systems
Recommended
Test Equipment
Recommended equipment for calibration is listed in Table 7-1. Other
t
est instruments may be used only if their specifications equal or
exceed the required characteristics.
Table 7-1, Recommended Test Equipment
Equipment Model No. Manufacturer
Oscilloscope (with jitter package)* LC684D LeCroy
Digital Multimeter (DMM) 2000 Keithley
Frequency Counter (Rubidium reference)*
2202R
Racal Instruments
Function Generator (with manual trigger)* WW2571A Tabor Electronics
Accessories BNC to BNC coaxial cables
BNC “Tconnector (2 BNC females with 1 BNC male)*
Adapter, dual banana jacks with BNC male
Feed-through termination, 50Ω ± 0.1%, 5W
* Required for Factory Calibration only.
Environment
Although Model 3152B can operate at ambient temperatures from 0°C
to 50°C, calibration should be performed under laboratory conditions
with an ambient temperature of 25°C ± 5°C and relati
ve humidity of
less than 80%.
Configuration
Calibration is performed with the 3152B installed in a VXIbus chassis.
Either a register-based or message-based controller may
be used,
and the procedure is the same for both options. T
he ArbConnection
software application must be installed and interfaced to the instrument.
Factory Calibration requires that the covers be removed during
calibration. Periodic Calibration does not require removal of the
covers, which should remain closed.
Periodic
Calibration
Calibration Interval
To ensure continued accuracy, it is recommended that periodic
calibration be performed at intervals of one year.