3152B User Manual Publication No. 980935 Rev. M
6-42
Performance Checks Astronics Test Systems
2. Connect the function generator to the 3152B TRIG IN connector
3. Configure the 3152B, as follows:
Auxiliary Function: Counter/Timer
Function: Totalize, Infinite
Trigger Level: 0 V
Test Procedure:
1. Press the Reset/Arm button on the 3152B to reset and arm
the totalize function.
2. Manually trigger the function generator and verify that the
3152B counter reading is 1,000,000 ±2.
Test Results Pass Fail