Publication No. 980935 Rev. M 3152B User Manual
Astronics Test Systems Performance Checks
6-41
2. Change the counter/timer function to Period Averaged
3. With the last function generator setting in Table 31, verify that
the period reading is 10.00000 µs ±50 ps
Test Results Pass Fail
Pulse Width
Equipment: Function Generator with at least 1 ppm accuracy
Preparation:
1. Configure the function generator as follows:
Frequency: As required by the test
Wave: Square
Duty Cycle: As required by the test
Amplitude 500 mV
2. Connect the function generator to the 3152B
TRIG IN connector
3. Configure the 3152B, as follows:
Auxiliary Function: Counter/Timer
Function: Pulse Width
Trigger Level: 0 V
Test Procedure:
1. Perform Pulse Width Accuracy tests using Table 6-37.
Table 6-37, Pulse Width Measurement Accuracy
Function Generator Setting
Error Limits
3152B
Counter Reading
Pass
Fail
Frequency
Duty Cycle
100 kHz
50 %
5.000 µs ±100 ns
100 kHz
70 %
7.07 µs ±100 ns
2. Change the counter/timer slope to Negative.
3. With the last function generator setting in Table 30, change the
function generator duty cycle to 70%.
4. Verify that the pulse width reading is 3.0 µs ±100 ns.
Test Results Pass Fail
Tot
alize, Infinite
Equipment: Function Generator with at least 1 ppm accuracy
Preparation:
1. Configure the function generator as follows:
Frequency: 100 MHz
Wave: Square
Amplitude 500 mV
Counted Burst: 1,000,000