3152B User Manual Publication No. 980935 Rev. M
6-40
Performance Checks Astronics Test Systems
Frequency
Equipment: Function Generator with at least 1 ppm accuracy
Preparation:
1. Configure the function generator as follows:
Frequency: As required by the test
Wave: Square
Amplitude 500 mV
2. Connect the function generator to the 3152B TRIG IN connector
.
3. Configure the 3152B, as follows:
Auxiliary Function: Counter/Timer
Function: Frequency
Trigger Level: 0 V
Test Procedure:
1.
Perform Frequency Measurement Accuracy tests using
Table 6-35.
Table 6-35, Frequency Measurement Accuracy
Function
Generator Setting
Error Limits
3152B
Counter Reading
Pass
Fail
1.000000 MHz
±2 Hz
100.0000 MHz
±100 Hz
Period, Period
Averaged
Equipment: Function Generator with at least 1 ppm accuracy
Preparation:
1. Configure the function generator as follows:
Frequency: As required by the test
Wave: Square
Amplitude 500 mV
2. Connect the function generator to the 3152B
TRIG IN connector
3. Configure the 3152B, as follows:
Auxiliary Function: Counter/Timer
Function: Period
Trigger Level: 0 V
Test Procedure:
1. Perform Period Accuracy tests using Table 6-36.
Table 6-36, Period Measurement Accuracy
Function
Generator Setting
Error Limits
3152B
Counter Reading
Pass
Fail
10 kHz
100.0 µs ±100 ns
100 kHz
10.00 µs ±100 ns