User manual

Table Of Contents
Publication No. 980935 Rev. M 3152B User Manual
Astronics Test Systems Performance Checks
6-39
Sync: On
Output: On
Test Procedure:
1. Verify Sweep operation on the oscilloscope as follows:
Waveform: Ramp up
Frequency: 1 kHz
Max Freq: 2 MHz
Min Freq: 1 MHz
Test Results Pass Fail
2. Move 3152B sweep marker position to 1.5 MHz and verify
marker position at the middle of the ramp.
Test Results Pass Fail
3. Reverse between Start and Stop frequencies and verify
oscilloscope reading as before except the ramp is down.
Test Results Pass Fail
4. Change sweep step to logarithmic and verify oscilloscope
exponential down waveform with properties as in 3 above.
Test Results Pass Fail
Auxiliary
Counter/Timer
Operation
This tests the operation of the auxiliary counter/timer function. Note
that when you select the counter/timer function all other 3152B
waveform generation is automatically purged and the instrument
operational mode is transformed to a stand-
alone counter/timer.
Waveform generation is resumed as soon as the counter/timer
function is turned off.