3152B User Manual Publication No. 980935 Rev. M
6-38
Performance Checks Astronics Test Systems
Time Base: 50 µs
Sampling Rate: 50 MS/s at least.
Trigger source: Channel 2, positive slope
Amplitude: 5 V/div
2. Connect the 3152B output to the oscilloscope input, channel 1.
3. Connect the 3152B SYNC to the oscilloscope input, channel 2.
4. Configure the 3152B controls as follows:
Waveform: Modulated
Modulation: Amplitude Hopping
Hop Mode: Fix
Dwell Time: 50 µs
Sync: On
Output: On
5. Using ArbConnection, open and prepare the Hop Table
composer and download the following table:
Amplitude
1
2
4
8
16
Test Procedure:
1. Verify 5 Hop operation on the oscilloscope as follows:
Waveform: Amplitude steps, 50 µs fixed dwell time
Min Amp: 1 V
Max Amp: 16 V
Period: 250
µ
s
Test Results Pass Fail
Sweep
Equipment: Oscilloscope
Preparation:
1. Configure the oscilloscope as follows:
Time Base: 0.2 ms
Sampling Rate: 50 MS/s at least.
Trace A View: Jitter, Type: FREQ, CLK.
Trigger source: Channel 2, positive slope
Amplitude: 1 V/div
2. Connect the 3152B output to the oscilloscope input, channel 1.
3. Connect the 3152B SYNC to the oscilloscope input, channel 2
4. Configure model 3152B controls as follows:
Waveform: Modulated
Modulation: Sweep
Start Frequency: 1 MHz
Stop Frequency: 2 MHz
Sweep Time: 1 ms
Sweep Type: Linear