Publication No. 980935 Rev. M 3152B User Manual
Astronics Test Systems Performance Checks
6-37
Test Results Pass Fail
Fix Dwell Time
Frequency Hop
ping
Equipment: Oscilloscope
Preparation:
1. Configure the oscilloscope as follows:
Time Base: 0.1 ms
Sampling Rate: 50 MS/s at least.
Trace A View: Jitter, Type: FREQ, CLK.
Trigger source: Channel 2, positive slope
Amplitude: 1 V/div
2. Connect the 3152B output to the oscilloscope input, channel 1.
3. Connect the 3152B SYNC to the oscilloscope input, channel 2.
4. Configure model 3152B controls as follows:
Waveform: Modulated
Modulation: Hop
Hop Mode: Fix
Dwell Time: 50 µs
Sync: On
Output: On
5. Using ArbConnection prepare, open the Hop Table composer
and download the following table:
Frequency
1.0e6
1.2e6
1.4e6
1.6e6
1.8e6
2.0e6
2.2e6
2.4e6
2.6e6
2.8e6
Test Procedure:
1. Verify Hop operation on the oscilloscope as follows:
Waveform: Frequency steps, fixed dwell time of 50 µs
Max Hop Freq: 2.8 MHz
Min Hop Freq: 1.0 MHz
Period: 500
µ
s
Test Results Pass Fail
Amplitude Hop
ping
Equipment: Oscilloscope
Preparation:
1. Configure the oscilloscope as follows: