User manual

Table Of Contents
3152B User Manual Publication No. 980935 Rev. M
6-36
Performance Checks Astronics Test Systems
Output: On
5. Using ArbConnection, prepare and download 10-step ASK list
with alternating “0” and “1”
Test Procedure:
1. Verify ASK operation on the oscilloscope as follows:
Waveform: Sine wave
Period: 0.1 ms
Amplitude: Every 0.1 ms alternates between 2 V and 4V
Test Results Pass Fail
Variable Dwell Time
Frequency Hop
ping
Equipment: Oscilloscope
Preparation:
1. Configure the oscilloscope as follows:
Time Base: 0.5 ms
Sampling Rate: 50 MS/s at least.
Trace A View: Jitter, Type: FREQ, CLK.
Trigger source: Channel 2, positive slope
Amplitude: 1 V/div
2. Connect the 3152B output to the oscilloscope input, channel 1
3. Connect the 3152B SYNC to the oscilloscope input, channel 2
4. Configure model 3152B controls as follows:
Waveform: Modulated
Modulation: Hop
Hop Mode: Variable
Sync: On
Output: On
5. Using ArbConnection prepare, open the Hop Table composer
and download the following table (both channels):
Frequency Dwell Time
1.0e6 50e-6
1.2e6 100e-6
1.4e6 150e-6
1.6e6 200e-6
1.8e6 250e-6
2.0e6 300e-6
2.2e6 350e-6
2.4e6 400e-6
2.6e6 450e-6
2.8e6 500e-6
Test Procedure:
1. Verify Hop operation on the oscilloscope as follows:
Waveform: Frequency steps, increasing dwell time from
50 µs to 500 µs
Max A: 2.8 MHz
Min A: 1.0 MHz
Period: 2750
µ
s