Publication No. 980935 Rev. M 3152B User Manual
Astronics Test Systems Performance Checks
6-31
3. Connect the 3152B SYNC to the oscilloscope input, channel
2.
4. Confgure the function generator as follows:
Frequency 1 kHz
Run Mode: Continuous
Waveform: Square wave.
Amplitude: 2 V Square
5. Connect the function generator output connector to the 3152B
TRIG IN connector.
6. Configure model 3152B controls as follows:
Waveform: Modulated
Modulation: FM
Modulation Run Mode: Burst
Burst: 5
Carrier Freq: 1 MHz
Mod Frequency: 10 kHz
Deviation: 500 kHz
Sync: On
Output: On
Test Procedure:
1. Verify Burst FM standard waveform operation on the
oscilloscope as follows:
Waveform: Burst of 5 Sine waveforms
Mod Frequency: 10 kHz
Burst Period: 1 ms
Max Deviation: 1.25 MHz
Min Deviation: 750 kHz
Test Results Pass Fail
Gated FM
- Standard
Waveforms
Equipment: Oscilloscope, function generator
Preparation:
1. Configure the oscilloscope as follows:
Time Base: 0.2 ms
Sampling Rate: 50 MS/s at least.
Trace A View: Jitter, Type: FREQ, CLK.
Trigger source: Channel 2, positive slope
Amplitude: 1 V/div
2. Connect the 3152B output to the oscilloscope input, channel 1.
3. Connect the 3152B SYNC to the oscilloscope input, channel 2.
4. Configure the function generator as follows:
Frequency 1 kHz
Run Mode: Continuous
Waveform: Square wave.
Amplitude: 2 V
5. Connect the function generator output connector to the 3152B
TRIG IN connector.
6. Configure model 3152B controls as follows: