Astronics Test Systems Performance Checks
6-29
on the FM modulation function only.
FM
- Standard
Waveforms
Equipment: Oscilloscope
Preparation:
1. Configure the oscilloscope as follows:
Time Base: 50 µs
Sampling Rate: 50 MS/s at least.
Trace A View: Jitter, Type: FREQ, CLK.
Trigger source: Channel 2, positive slope
Amplitude: 1 V/div
2. Connect the 3152B output to the oscilloscope input, channel 1.
3. Connect the 3152B SYNC to the oscilloscope input, channel 2.
4. Configure model 3152B controls as follows:
Waveform: Modulated
Modulation: FM
Carrier Freq: 1 MHz
Mod Frequency: 10 kHz
Deviation: 500 kHz
Sync: On
Output: On
Test Procedure:
1. Verify FM operation on the oscilloscope as follows:
Waveform: Sine
Frequency: 10 kHz
Max F: 1.25 MHz
Min F: 750 kHz
Test Results Pass Fail
2. Modify the 3152B modulating waveform to triangle, then square,
and then ramp to verify all standard FM waveforms.
Test Results Pass Fail
3. Move the 3152B marker position to 1.25MHz and verify the
marker position.