User manual

Table Of Contents
3152B User Manual Publication No. 980935 Rev. M
6-26
Performance Checks Astronics Test Systems
Table 6-32, PLL Tests Frequency
Function Generator
Frequency Setting
3152B
Frequency
Setting
Low
Limit
Counter
Ratio
Reading
High
Limit
Pass
Fail
500 Hz
100 kHz
0.99
1.01
5 kHz
100 kHz
0.99
1.01
50 kHz
100 kHz
0.99
1.01
500 kHz
1 MHz
0.99
1.01
5 MHz
10 MHz
0.99
1.01
10 MHz
10 MHz
0.99
1.01
PLL Checks
Phase
Offset
Equipment: Counter, function generator
Preparation:
1. Configure the function generator as follows:
Waveform: Square
Amplitude: 1 Vp-p
Output: On
Frequency: As required (see Table 6-32)
2. Referring to Figure 6-1 (above), connect the function generator
output to a “T” connector at the counter CH A input. Connect
the other side of the “T” connector to the 3152B TRIG/PLL IN
connector. Use a 50 feed-through terminator at the 3152B
TRIG/PLL IN connector.
3. Configure the counter as follows:
Function: φ AB
CH A Input: High impedance
CH B Input: 50 Ω
4. Connect the 3152B output to the counter CH B input.
5. Configure the 3152B as follows:
Function Mode: Arbitrary
Wavelength: 200 points
Output: On
PLL: On
Trigger Level: 0 V
Phase Offset: As required (see Table 6-33)
Test Procedure:
1. For each test step (row) in Table 6-
33, set the function
generator frequency and 3152B phase as indicated.
2. Read and record the phase reading displayed by the counter
and verify that it is within the low and high limits given in
Table 6-33.