Publication No. 980935 Rev. M 3152B User Manual
Astronics Test Systems Performance Checks
6-25
Function: Freq A/B
CH A Input: High impedance
CH B Input: 50
4. Connect the 3152B output to the counter CH B input.
5. Configure the 3152B as follows:
Impedance: 1 MΩ
Waveform: Sine
Output: On
PLL: On
Trigger Level: 0 V
Frequency: As required (see Table 6-32)
Test Procedure:
1. For each test step (row) in Table 6-
32, set the function
generator and 3152B frequencies as indicated.
2. Read and record the frequency ratio displayed by the counter
and verify that it is within the low and high limits given in
Table 6-32.
Figure 6-1, PLL Tests Signal Connections