User manual

Table Of Contents
3152B User Manual Publication No. 980935 Rev. M
6-24
Performance Checks Astronics Test Systems
Test Results Pass Fail
SYNC Output
Zero
Equipment: Oscilloscope
Preparation:
1. Configure the oscilloscope as follows:
Time Base: As required by the test
Amplitude: 2 V/div
2. Connect the 3152B output to oscilloscope input (1)
3. Connect the 3152B SYNC to oscilloscope input (2)
4. Configure model 3152B channel as follows:
Waveform: Sinc
‘0’ Crossing 10
Output: On
Sync Output: On
Sync Source: Zero
Test Procedure:
1. Observe that the SYNC output has 10 sinc cycles for each
“0” crossing level.
Test Results Pass
Fail
PLL Operation
This tests the operation of the PLL function. Model 3152B locks
automatically to an external trigger source. The frequency and the
start phase of the external signals are applied to the 3152B TRIG/PLL
input. After lock (as indicated by the LOCK LED)
, the start phase of
the 3152B can be modified from -180° to 180°.
PLL Checks
Frequency Lock
Equipment: Counter, function generator
Preparation:
1. Configure the function generator as follows:
Waveform: Square
Amplitude: 1 Vp-p
Output: On
Frequency: As required (see Table 6-32)
2. Referring to Figure 6-1, connect the function generator output
to a “T” connector at the counter CH A input. Connect the other
side of the “T” connector to the 3152B TRIG/PLL IN connector.
Use a 50 fee`d-through terminator at the 3152B TRIG/PLL IN
connector.
3. Configure the counter as follows: