Publication No. 980935 Rev. M 3152B User Manual
Astronics Test Systems Performance Checks
6-23
1. From ArbConnection, click on the Manual Trigger button and
verify that the trace on the oscilloscope shows synchronization
pulses having 10 µs pulse widths. Verify that the SYNC pulse is
high for the duration of the burst.
Test Results Pass Fail
SYNC
Output - HCL
Equipment: Oscilloscope
Preparation:
1. Configure the oscilloscope as follows:
Time Base: As required by the test
Amplitude: 2 V/div
2. Connect the 3152B output to the oscilloscope input (1)
3. Connect the 3152B SYNC to the oscilloscope input (2)
4. Configure model 3152B channel as follows:
Waveform: Sine
Output: On
Sync Output: On
Sync Source: HCL
Test Procedure:
1. Observe that the SYNC output has a 50% duty cycle.
Test Results Pass Fail
SYNC
Output - Pulse
Equipment: Oscilloscope
Preparation:
1. Configure the oscilloscope as follows:
Time Base: As required by the test
Amplitude: 2 V/div
2. Connect the 3152B output to the oscilloscope input (1)
3. Connect the 3152B SYNC to the oscilloscope input (2)
4. Configure model 3152B channel as follows:
Waveform: Sine
Output: On
Sync Output: On
Sync Source: Pulse
Sync Width: 8
Sync Position: 48
Test Procedure:
1. Observe that the SYNC output has a width of 8 sample clock
cycles and that its position is shifted by 48 sample clock
periods.