User manual

Table Of Contents
3152B User Manual Publication No. 980935 Rev. M
6-22
Performance Checks Astronics Test Systems
Test Results Pass Fail
SYNC Output
O
peration
This tests the operation of the SYNC output. There are two
synchronous output are being tested BIT and LCOM. BIT normally
operates with standard and arbitrary waveforms and LCOM is
associated with sequenced and burst outputs. The sync output has a
fixed TTL level amplitude into an open circuit.
SYNC
Output - BIT
Equipment: Oscilloscope
Preparation:
1. Configure the oscilloscope as follows:
Time Base: As required by the test
Amplitude: 2 V/div
2. Connect 3152B SYNC to the oscilloscope input.
3. Configure model 3152B as follows:
Waveform: Sine
Output: On
Sync Output: On
Test Procedure:
1. Verify that the
trace on the oscilloscope shows synchronization
pulses at 1 µs intervals.
Test Results Pass Fail
SYNC
Output -
LCOM
Equipment: Oscilloscope
Preparation:
1. Configure the oscilloscope as follows:
Time Base: As required by the test
Amplitude: 2 V/div
2. Connect the 3152B output to the oscilloscope input (1).
3. Connect the 3152B SYNC to the oscilloscope input (2).
4. Configure model 3152B channel as follows:
Waveform: Sine
Run Mode: Burst
Burst Count: 10
Re-trigger On
Re-trig period: 10 µs
Output: On
Sync Output: On
Test Procedure: