Publication No. 980935 Rev. M 3152B User Manual
Astronics Test Systems Performance Checks
6-21
5. Connect the 3152B to the Oscilloscope input.
6. Configure the 3152B as follows:
SCLK 100 MS/s
Waveform: Sequence
Seq. Advance: Step
Amplitude: 2 V
Trigger Source: External
Output: On
7. Using ArbConnection prepare and download the following
waveform:
Segment 1: Sine, 1000 points
Segment 2: Triangle, 1000 points
Segment 3: Square, 1000 points
Segment 4: Sinc, 1000 points
Segment 5: Gaussian Pulse, 1000 points
8. Using ArbConnection, build and download the following
sequence table:
Step 1: Segment 1, loop 1
Step 2: Segment 2, loop 1
Step 3: Segment 3, loop 1
Step 4: Segment 4, loop 1
Step 5: Segment 5, loop 1
Test Procedure
1. From ArbConnection, click on the Manual Trigger and
observe that the waveforms advance through the sequence
table repeatedly.
Test Results Pass Fail
Note
Leave the same setup for the next test
Single Advance
Equipment: Oscilloscope, function generator
Preparation: (Same preparation as for previous step, except change
mode to single sequence advance)
1. Change Oscilloscope configuration to single.
Test Procedure
1. From ArbConnection, click on Manual Trigger and observe
that one waveform cycle advances through the sequence
table repeatedly with each external trigger signal. Note that
you might need to reset the single-shot capture mode on the
oscilloscope for each trigger advance.