3152B User Manual Publication No. 980935 Rev. M
6-20
Performance Checks Astronics Test Systems
Automatic Advance
Equipment: Counter
Preparation:
1. Configure the Counter as follows:
Function: TOTB Measurement
2. Connect the counter channel B to the 3152B output
3. Configure the 3152B as follows:
SCLK 100 MS/s
Waveform: Sequence
Run Mode: Trigger
Trigger Source: BUS
Amplitude: 2 V
Output: On
4. Using ArbConnection prepare and download the following
waveform:
Segments: 1 to 5
Wavelength: 128 points
Waveform: 1 cycle square
5. Using ArbConnection, build and download the following
sequence table:
Step 1: Segment 1, loop 100,000
Step 2: Segment 2, loop 100,000
Step 3: Segment 3, loop 100,000
Step 4: Segment 4, loop 100,000
Step 5: Segment 5, loop 100,000
Test Procedure
1. From ArbConnection, click on the Manual Trigger button and
observe that counter reading is 500,000 counts. Reset
counter and repeat the test a few times. Every time the
counter reading should be 500,000 counts exactly.
Test Results Pass Fail
Step Advance
Equipment: Oscilloscope, function generator
Preparation:
1. Configure the Oscilloscope as follows:
Termination: 50
, 20 dB feedthrough attenuator at the
oscilloscope input
Setup: As required for the test
2. Connect the 3152B output to the oscilloscope input.
3. Configure the function generator as follows:
Frequency 10 kHz
Run Mode: Triggered
Waveform: Square wave.
Amplitude: Adjust for TTL level on 50
4. Connect the function generator output to the 3152B TRIG IN
connector.