User manual

Table Of Contents
Publication No. 980935 Rev. M 3152B User Manual
Astronics Test Systems Performance Checks
6-19
Termination: 50 , 20 dB feedthrough attenuator at the
oscilloscope input
Setup: As required for the test
2. Connect the 3152B output to the oscilloscope input
3. Configure the 3152B as follows:
Frequency: 2 MHz
Run Mode: Trigger
Trigger Source: As specified in Table 6-31
Waveform: Sine wave
Amplitude: 2 V
Output: On
4. Configure the auxiliary 3152B as follows:
Frequency: 1 MHz
Waveform: Sine wave
Run Mode: Continuous
Trigger Output: As specified in Table 6-31
Output: On
Test Procedure
1. For each test step (row) in Table 6-31, set the “Trigger
Outputand “Trigger Sourceas specified. Verify that the
3152B generates a single-cycle, 2 MHz sine wave at 1 µs
intervals.
Table 6-31, Trigger Source Tests
Auxiliary 3152B
Trigger Output
3152B
Trigger Source"
Oscilloscope
Reading
Pass
Fail
TTLT0 ON
TTLT0
Single-cycle sine wave at
1 µs intervals
TTLT0 OFF
TTLT1 ON
TTLT1
Single-cycle sine wave at
1 µs intervals
TTLT1 OFF
TTLT2 ON
TTLT2
Single-cycle sine wave at
1 µs intervals
TTLT2 OFF
TTLT3 ON
TTLT3
Single-cycle sine wave at
1 µs intervals
TTLT3 OFF
TTLT4 ON
TTLT4
Single-cycle sine wave at
1 µs intervals
TTLT4 OFF
TTLT5 ON
TTLT5
Single-cycle sine wave at
1 µs intervals
TTLT5 OFF
TTLT6 ON
TTLT6
Single-cycle sine wave at
1 µs intervals
TTLT6 OFF
TTLT7 ON
TTLT7
Single-cycle sine wave at
1 µs intervals
Sequence
operation
This tests the operation of the sequence generators. This also checks
the various sequence advance options.