Publication No. 980935 Rev. M 3152B User Manual
Astronics Test Systems Performance Checks
6-15
Table 6-27, Sine Wave Flatness Test, DDS Output at 10 V
3152B Sine
Frequency
Error Limits
Oscilloscope Reading
Pass
Fail
1 kHz
6 Divisions
Reference
X
X
1 MHz
6 ±0.3 Divisions
10 MHz
6 ±0.3 Divisions
50 MHz
6 ±0.9 Divisions
Trigger O
perational
Characteristics
This tests the operation of the trigger circuit including
tests for the
triggered, gated,
and counted burst run modes. It also tests the
operation of the triggered advance options, the delayed trigger, the re-
trigger functions, as well as the trigger input level and slope sensitivity.
Trigger, Gate, and
Burst
Characteristics
Equipment: Oscilloscope, function generator, counter
Preparation:
1. Configure the Oscilloscope as follows:
Termination: 50 , 20
dB feedthrough attenuator at the
oscilloscope input
Setup: As required for the test
2. Configure the counter as follows:
Function: TOT B
Trigger Level: 100 mV
3. Connect 3152B output to the oscilloscope input
4. Configure the function generator as follows:
Frequency 1 MHz
Run Mode: As required by the test
Wave: 2 V Square
5. Connect the function generator output to the 3152B TRIG IN
connector
6. Configure the 3152B as follows:
Frequency: 25 MHz
Waveform: Sine wave
Burst Count: 1e6 counts
Amplitude: 1 V
Trigger Source: External
Output: On
Test Procedure
1. Perform trigger and gate tests using Table 6-28.
Table 6-28, Trigger, Gate, and Burst Characteristics
3152B Run
External Trigger
Oscilloscope/Counter
Mode
Pulse
Reading
Pass
Fail
Triggered
1 MHz, Continuous
Triggered waveform
GatedTransition
1 MHz, Continuous
Gated by transition Waveform
GatedLevel
1 MHz, Continuous
Gated by level Waveform
Burst
Single shot
Burst, 1e6 waveforms