3152B User Manual Publication No. 980935 Rev. M
6-14
Performance Checks Astronics Test Systems
3. Change amplitude to 10V and adjust the vertical controls of
the Oscilloscope to get 6 division of display. Perform sine
wave flatness, DAC waveforms tests using Table 6-25.
Table 6-25, Sine Wave Flatness, DAC Output Test at 10 V
3152B Sine
Frequency
Error Limits
Oscilloscope Reading
Pass
Fail
1 kHz
6 Divisions
Reference
X
X
1 MHz
6 ±0.3 Divisions
10 MHz
6 ±0.3 Divisions
50 MHz
6 ±0.6 Divisions
100 MHz
6 ±0.6 Divisions
Sine Wave
Flatness,
DDS Output
Equipment: Oscilloscope
Preparation:
1. Configure the Oscilloscope follows:
Termination: 50
, 20 dB feedthrough attenuator at the
oscilloscope input
Setup: As required for the test
2. Connect 3152B output to the oscilloscope input
3. Configure the 3152B as follows:
Waveform: Modulated
Modulation: OFF
Amplitude: 5 V
Output: On
CW Frequency: Initially, 1 kHz then, as required by the test
Test Procedure
1. Adjust the vertical controls of the Oscill
oscope to get 6 division of
display.
2. Perform Sine flatness, DDS waveforms tests using Table 6-26.
Table 6-26, Sine Wave Flatness Test, DDS Outputat 5 V
3152B Sine
Frequency
Error Limits
Oscilloscope Reading
Pass
Fail
1 kHz
6 Divisions
Reference
X
X
1 MHz
6 ±0.2 Divisions
10 MHz
6 ±0.3 Divisions
50 MHz
6 ±0.6 Divisions
3. Change amplitude to 10 V and adjust the vertical controls of the
Oscilloscope to get 6 division of display
. Perform sine wave
flatness, DAC waveforms tests using Table 6-27.