3152B User Manual Publication No. 980935 Rev. M
6-10
Performance Checks Astronics Test Systems
1. Perform Offset Accuracy tests using Table 6-16.
Table 6-16, Offset Accuracy, DDS Outputat 6 V
3152B Offset
Setting
Error Limits
DMM Reading
Pass
Fail
0.000 V
0 ±130 mV
2. Modify 3152B Amplitude setting to 1 V and offset setting to 0 V.
3. Continue the Offset tests using Table 6-17.
Table 6-17, Offset Accuracy, DDS Output at 1 V
3152B Offset
Setting
Error Limits
DMM Reading
Pass
Fail
0.000 V
0 ±30 mV
Square
Wave
Characteristics
This tests the characteristics of the square waveform including
transition times, ringing, and overshoot.
Square
Wave
Checks
Equipment: Oscilloscope, 50 , 20 dB attenuator feed through
Preparation:
1. Configure the Oscilloscope follows:
Termination: 50
, 20 dB attenuator feed through at the
oscilloscope input
Setup: As required for the test
2. Connect 3152B output to the oscilloscope input
3. Configure the 3152B as follows:
Frequency: 1 MHz
Waveform: Square wave
Amplitude: 6 V
Output: On
Test Procedure
1. Perform Square Wave Characteristics tests using Table 6-18.
Table 6-18, Square Wave Characteristics at 6 V
Parameter
Tested
Error Limits
Oscilloscope Reading
Pass
Fail
Rise/Fall Time
<5 ns
Ringing
<6 % + 10 mV
Over/undershoot
<6 % + 10 mV