User manual

Table Of Contents
Publication No. 980935 Rev. M 3152B User Manual
Astronics Test Systems Performance Checks
6-9
Output: On
Offset: As specified in Table 6-13
Test Procedure
1. Perform Offset Accuracy tests using Table 6-13.
Table 6-13, Offset Accuracy, DAC Outputat 20 mV
3152B Offset
Setting
Error Limits
DMM Reading
Pass
Fail
+4.000 V
8.000 V ±90 mV
+1.500 V
3.000 V ±40 mV
0.000 V
0 V ±40 mV
-1.500 V
-3.000 V ±40 mV
-4.000 V
-8.000 V ±90 mV
2. Modify 3152B Amplitude setting to 6 V and offset setting to 0 V.
3. Continue the Offset tests using Table 6-14.
Table 6-14, Offset Accuracy, DAC Output at 6 V
3152B Offset
Setting
Error Limits
DMM Reading
Pass
Fail
0.000 V
0 ±130 mV
4. Modify 3152B Amplitude setting to 1 V and offset setting to 0 V.
5. Continue the Offset tests using Table 6-15.
Table 6-15, Offset Accuracy, DAC Output at 1 V
3152B Offset
Setting
Error Limits
DMM Reading
Pass
Fail
0.000 V
0 ±30 mV
Offset Accuracy,
DDS Output
, 1 M
Load Impedance
Equipment: DMM
Preparation:
4. Configure the DMM as follows:
Termination: Open circuit at the DMM input
Function: DCV
5. Connect 3152B output to the DMM input
6. Configure the 3152B as follows:
outp:load:cal: 1M
Waveform: Modulated
Modulation: OFF
CW Frequency: 1 MHz
Amplitude: 6 V
Output: On
Test Procedure