User manual

Table Of Contents
3152B User Manual Publication No. 980935 Rev. M
6-8
Performance Checks Astronics Test Systems
Table 6-10, Offset Accuracy, DAC Output at 1 V
3152B Offset
Setting
Error Limits
DMM Reading
Pass
Fail
0.000 V
0 ±10 mV
Offset Accuracy,
DDS Output
, 50
Load Impedance
Equipment: DMM
Preparation:
1. Configure the DMM as follows:
Termination: 50 feedthrough at the DMM input
Function: DCV
2. Connect 3152B output to the DMM input
3. Configure the 3152B as follows:
Waveform: Modulated
Modulation: OFF
CW Frequency: 1 MHz
Amplitude: 6 V
Output: On
Test Procedure
1. Perform Offset Accuracy tests using Table 6-11.
Table 6-11, Offset Accuracy, DDS Outputat 6 V
3152B Offset
Setting
Error Limits
DMM Reading
Pass
Fail
0.000 V
0 ±65 mV
1. Modify 3152B Amplitude setting to 1 V and offset setting to 0 V.
2. Continue the Offset tests using Table 6-12.
Table 6-12, Offset Accuracy, DDS Output at 1 V
3152B Offset
Setting
Error Limits
DMM Reading
Pass
Fail
0.000 V
0 ±15 mV
Offset Accuracy,
DAC Output
, 1 M
Load Impedance
Equipment: DMM
Preparation:
1. Configure the DMM as follows:
Termination: Open circuit at the DMM input
Function: DCV
2. Connect 3152B output to the DMM input
3. Configure the 3152B as follows:
outp:load:cal: 1M
Frequency: 1 MHz
Amplitude: 20 mV