User manual

Table Of Contents
Publication No. 980935 Rev. M 3152B User Manual
Astronics Test Systems Performance Checks
6-7
Offset Accuracy
The offset accuracy test checks the accuracy of the offset generator.
The o
ffset path is checked for both the DAC route (arbitrary and
standard waveforms) a
nd the DDS route (CW and modulated
waveforms). The accuracy is also tested separately for 50 and for 1
M load impedances.
Offset Accuracy,
DAC Output
, 50
Load Impedance
Equipment: DMM
Preparation:
1. Configure the DMM as follows:
Termination: 50 feedthrough at the DMM input
Function: DCV
2. Connect 3152B output to the DMM input
3. Configure the 3152B as follows:
Frequency: 1 MHz
Amplitude: 20 mV
Output: On
Offset: As specified in Table 6-8
Test Procedure
1. Perform Offset Accuracy tests using Table 6-8.
Table 6-8, Offset Accuracy, DAC Outputat 20 mV
3152B Offset
Setting
Error Limits
DMM Reading
Pass
Fail
+4.000 V
4.000 V ±45 mV
+1.500 V
1.500 V ±20 mV
0.000 V
0 V ±20 mV
-1.500 V
-1.500 V ±20 mV
-4.000 V
-4.000 V ±45 mV
1. Modify 3152B Amplitude setting to 6 V and offset setting to 0 V.
2. Continue the Offset tests using Table 6-9.
Table 6-9, Offset Accuracy, DAC Output at 6 V
3152B Offset
Setting
Error Limits
DMM Reading
Pass
Fail
0.000 V
0 ±65 mV
1. Modify 3152B Amplitude setting to 1 V and offset setting to 0 V.
2. Continue the Offset tests using Table 6-10.