User manual

Table Of Contents
3152B User Manual Publication No. 980935 Rev. M
6-6
Performance Checks Astronics Test Systems
outp:load:cal: 1M
Frequency: 1 kHz
Output: On
Amplitude: As specified in Table 6-6
Test Procedure
1. Perform amplitude accuracy tests using Table 6-6.
Table 6-6, Amplitude Accuracy, DAC Output
3152B Amplitude
Setting
Error Limits
DMM Reading
Pass
Fail
10.00 V
7.071 V ± 170 mV
5.000 V
3.535 V ± 135 mV
2.500 V
1.767 V ± 117 mV
1.000 V
707.1 mV ± 47 mV
500 mV
353.5 mV ± 43 mV
100.0 mV
70.71 mV ± 10.7 mV
50.00 mV
35.35 mV ± 10.3 mV
Amplitude Accuracy,
DDS Output
, 1 M
Load Impedance
Equipment: DMM
Preparation:
1. Configure the DMM as follows:
Termination: Open circuit at the DMM input
Function: ACV
2. Connect 3152B output to the DMM input
3. Configure the 3152B as follows:
outp:load:cal: 1M
Waveform: Modulated
Modulation: OFF
CW Frequency: 1 kHz
Output: On
Amplitude: As specified in Table 6-7
Test Procedure
1. Perform amplitude Accuracy tests using Table 6-7.
Table 6-7, Amplitude Accuracy, DDS Output
3152B Amplitude
Setting
Error Limits
DMM Reading
Pass
Fail
10.00 V
7.071 V ± 170 mV
5.000 V
3.535 V ± 135 mV
2.500 V
1.767 V ± 117 mV
1.000 V
707.1 mV ± 47 mV
500 mV
353.5 mV ± 43 mV
100.0 mV
70.71 mV ± 10.7 mV
50.00 mV
35.35 mV ± 10.3 mV