User manual

Table Of Contents
Publication No. 980935 Rev. M 3152B User Manual
Astronics Test Systems Performance Checks
6-5
Table 6-4, Amplitude Accuracy, DAC Output
3152B Amplitude
Setting
Error Limits
DMM Reading
Pass
Fail
10.00 V
3.534 V ± 85 mV
5.000 V
1.767 V ± 67 mV
2.500 V
883.0 mV ± 59 mV
1.000 V
353.4 mV ± 23 mV
500 mV
176.7 mV ± 22 mV
100.0 mV
35.34 mV ± 5.4 mV
50.00 mV
17.67 mV ± 5.2 mV
Amplitude Accuracy,
DDS Output
, 50
Load Impedance
Equipment: DMM
Preparation:
1. Configure the DMM as follows:
Termination: 50 feedthrough at the DMM input
Function: ACV
2. Connect 3152B output to the DMM input.
3. Configure the 3152B as follows:
Waveform: Modulated
Modulation: OFF
CW Frequency: 1 kHz
Output: On
Amplitude: As specified in Table 6-5
Test Procedure
1. Perform amplitude Accuracy tests using Table 6-5.
Table 6-5, Amplitude Accuracy, DDS Output
3152B Amplitude
Setting
Error Limits
DMM Reading
Pass
Fail
10.00 V
3.534 V ± 85 mV
5.000 V
1.767 V ± 67 mV
2.500 V
883 mV ± 59 mV
1.000 V
353.4 mV ± 23 mV
500 mV
176.7 mV ± 22 mV
100.0 mV
35.34 mV ± 5.4 mV
50.00 mV
17.67 mV ± 5.2 mV
Amplitude Accuracy,
DAC Output
, 1 M
Load Impedance
Equipment: DMM
Preparation:
1. Configure the DMM as follows:
Termination: Open circuit at the DMM input
Function: ACV
2. Connect 3152B Channel to the DMM input.
3. Configure the 3152B as follows: