3152B User Manual Publication No. 980935 Rev. M
6-4
Performance Checks Astronics Test Systems
Frequency
Accuracy,
External
10
MHz Reference
Equipment: 10 MHz reference (at least 0.1ppm), Counter
Preparation:
1. Leave counter setting and 3152B connections as in previous
test.
2. Connect the 10 MHz reference oscillator to the 3152B 10 MHz
reference input.
3. Configure the 3152B as follows:
10 MHz: External
Waveform: Square wave
Amplitude: 2 V
Output: On
Frequency: As specified in Table 6-3
Test Procedure
1. Perform frequency accuracy tests using Table 6-3.
Table 6-3, Frequency Accuracy Using External 10 MHz Reference
3152B Setting
Error Limits
Counter Reading
Pass
Fail
10.000000000 MHz
±1 Hz
50.000000000 MHz
±5 Hz
Amplitude
Accuracy
The Amplitude Accuracy test verifies
the accuracy of the output
amplifier and attenuators. The amplitude path is checked for both the
DAC route (arbitrary and standard waveforms) and the DDS route
(CW and modulated waveforms). The accurac
y is also tested
separately for 50 and for 1 M load impedances.
Amplitude Accuracy,
DAC Output
, 50
Load Impedance
Equipment: DMM
Preparation:
1. Configure the DMM as follows:
Termination: 50 feedthrough at the DMM input
Function: ACV
2. Connect 3152B Channel to the DMM input.
3. Configure the 3152B as follows:
Frequency: 1 kHz
Output: On
Amplitude: As specified in Table 6-4
Test Procedure
1. Perform amplitude accuracy tests using Table 6-4.