User manual

Table Of Contents
3152B User Manual Publication No. 980935 Rev. M
5-62 Programming Reference Astronics Test Systems
<tre>
0 to 99.99
Numeric
10
Programs the trailing edge transition time of the
standard pulse waveform in units of percent.
Response
The
3152B returns the current frequency, amplitude, offset, delay time, pulse width and leading and trailing
edges settings similar to the following example: 1e6,5,0,10,10,10,10.
APPLy:RAMP<freq>,<ampl>,<offs>,<del>,<lee>,<tre>(?)
HS Syntax:
0305F<float>,<float,<float>,<float,<float>,<float>(?)
Description
This command
changes the waveform function to standard ramp and programs
the frequency, amplitude,
offset, delay time, pulse width
,
and leading and trailing edges simultaneously. This command affects the
output regardless of the current output function. For example, if
the 3152B is in FM mode, the 3152B will st
op
generating FM and
will revert to the standard ramp waveform with parametric values
as specified by this
command.
Parameters
Name
Range
Type
Default
Description
<freq>
10e
-3 to
100e6
Numeric
1e6
Programs the frequency of the standard ramp waveform
in units of Hz.
<ampl>
10e
-3 to
16
Numeric
5
Programs the amplitude of the standard ramp waveform
in units of
Volts.
<offs>
-
7.992 to
7.992
Numeric
0
Programs the offset of the standard ramp waveform in
units of
Volts.
<del>
0 to
99.99
Numeric
10
Programs the delay time of the standard ramp
waveform in units of percent.
<lee>
0 to 99.99
Numeric
10
Programs the leading edge transition time of the
standard ramp waveform in units of percent.
<tre>
0 to 99.99
Numeric
10
Programs the trailing edge transition time of the
standard ramp waveform in units of percent.
Response
The
3152B returns the current
frequency, amplitude, offset, delay time, pulse width and leading and trailing
edges settings similar to the following example: 1e
6,5,0,10,10,10.
APPLy:SINC <freq>,<ampl>,<offs>,<cycles>(?)
HS Syntax:
0310D<float>,<float,<float>,<integer>(?)
Description
This command
changes the waveform function to standard sinc and programs
the frequency, amplitude,
offset
, and “0” crossing cyc
les simultaneously. This command affects the output regardless of the current
output function. For example, if
the 3152B is in FM mode, the 3152B will stop generating FM and will revert to
the
standard sinc waveform with parametric values as specified by this command.
Parameters
Name
Range
Type
Default
Description
<freq>
10e
-3 to
100e6
Numeric
1e6
Programs the frequency of the standard sinc waveform
in units of Hz.