User manual

Table Of Contents
Publication No. 980935 Rev. M 3152B User Manual
Astronics Test Systems Programming Reference 5-61
APPLy:SQUare <freq>,<ampl>,<offs>,<dcycle>(?)
HS Syntax:
0303D<float>,<float,<float>,<float>(?)
Description
This command
changes the waveform function to standard square and programs
the frequency, amplitude,
offset,
and
duty cycle simultaneously. This command affects the output regardless of the current output
function. For example, if
the 3152B is in FM mode, the 3152B will stop generating FM and will revert to
the
standard
square waveform with parametric values as specified by this command.
Parameters
Name
Range
Type
Default
Description
<freq>
10e
-3 to
100e6
Numeric
1e6
Programs the frequency of the standard square
waveform in units of Hz.
<ampl>
10e
-3 to
16
Numeric
5
Programs the amplitude of the standard square
waveform in units of
Volts.
<offs>
-
7.992 to
7.992
Numeric
0
Programs the offset of the standard square waveform in
units of
Volts.
<dcycle>
0 to 99.99
Numeric
50
Programs the duty cycle of the standard square
waveform in units of percent.
Response
The
3152B returns the current frequency, amplitude, offset, and duty cycle settings as in
the following
example: 1e6,5,0,50.
APPLy:PULSe <freq>,<ampl>,<offs>,<del>,<wid>,<lee>,<tre>(?)
HS Syntax:
0304G<float>,<float,<float>,<float>,<float,<float>,<float>(?)
Description
This command
changes the waveform function to standard pulse and programs
the frequency, amplitude,
offset, delay
time, pulse width and leading and trailing edges simultaneously. This command affects the output
regardless of the current output function. For example, if you generate arbitrary waveforms, the 3152B will
stop generating arbitrary waveforms, will revert to the standard pulse waveform and will update the pulse
parameters, as specified by this command.
Parameters
Name
Range
Type
Default
Description
<freq>
10e
-3 to
100e6
Numeric
1e6
Programs the frequency of the standard pulse
waveform in units of Hz.
<a
mpl>
10e
-3 to
16
Numeric
5
Programs the amplitude of the standard pulse waveform
in units of
Volts.
<offs>
-
7.992 to
7.992
Numeric
0
Programs the offset of the standard pulse waveform in
units of
Volts.
<del>
0 to 99.99
Numeric
10
Programs the delay t
ime of the standard pulse
waveform in units of percent.
<wid>
0 to 99.99
Numeric
10
Programs the pulse width of the standard pulse
waveform in units of percent.
<lee>
0 to 99.99
Numeric
10
Programs the leading edge transition time of the
standard pulse
waveform in units of percent.