User manual

Table Of Contents
Publication No. 980935 Rev. M 3152B User Manual
Astronics Test Systems Programming Reference 5-45
HS Syntax: 0600A<0|1|2|3|4|5|6|7|8|9>(?)
Description
Defines the standard waveform shape to be output by the 3152B.
Parameters
Name
Type
Default
Description
SINusoid
Discrete
SIN
Selects the built
-in sine waveform.
TRIangle
Discrete
Selects the built
-in triangular waveform.
SQUare
Discrete
Selects the built
-in square waveform.
SPULse
Discrete
Selects the built
-in pulse waveform.
RAMP
Discrete
Selects the built
-in ramp waveform.
SINC
Discrete
Selects the built
-in sinc waveform.
EXPonential
Discrete
Selects the built
-in exponential waveform.
GAUSsian
Discrete
Selects the built
-in gaussian waveform.
DC
Dis
crete
Selects the built
-in DC waveform.
NOISe
Discrete
Selects the built
-in noise waveform.
Response
The
3152B returns SIN, TRI, SQU, SPUL, RAMP, SINC, GAUS, EXP, DC, or NOIS depending on the current
3152B setting.
SINusoid:PHASe <phase>(?)
HS Synta
x: 0700A<float>(?)
Description
Programs the start phase for the standard sine waveform. This command has no affect on modulated
waveforms.
Parameters
Name
Range
Type
Default
Description
<phase>
0 to 360
Numeric
0
Programs the start phase parameter in
units of degrees.
Response
The 3152B returns the current start phase value.
SINusoid:POWer <power>(?)
HS Syntax: 0701A<integer>(?)
Description
Programs power for the sine
x
waveform. This command has no affect on modulated waveforms.
Parameters
Name
Range
Type
Default
Description
<power>
1 to 9
Numeric
1
Programs the power that the sine function will be raised
to. Note that even powers will provide waveforms that
are positive (all points >= 0) whereas odd powers will
provide functions that are positive or negative 50% of