3152B User Manual Publication No. 980935 Rev. M
5-44 Programming Reference Astronics Test Systems
Standard
Waveform
Control
Commands
This group is used to control the standard waveforms and their
respective parameters. There is an array of standard waveforms that
could be used without the need to download waveform
values
to the
instrument. You can also modify the parameters for each waveform to
a shape suitable
to your application.
Factory defaults after *RST are shown in the Default column.
Parameter
range and low and high limits are listed, where applicable.
Table 5-5, Standard Waveform Control Commands Summary
Keyword
Parameter Form
Default
3152A
HS
[:SOURce]
:SHAPe SINusoid | TRIangle | SQUare | PULSe | RAMP | SINC |
GAUSsian | EXPonential | NOISe | DC
SIN
0600A
:SINusoid
:PHASe
0 to 360
0
0700A
:POWer
1 to 9
1
0701A
:TRIangle
:PHASe
0 to 360
0
0710A
:POWer
1 to 9
1
0711A
:SQUare
:DCYCle
0 to 99.99
50
0720A
:PULSe
:DELay
0 to 99.999
10
0730A
:WIDTh
0 to 99.999
10
0731A
:TRANsition
[:LEADing]
0 to 99.999
10
0732A
:TRAiling
0 to 99.999
10
0733A
:RAMP
:DELay
0 to 99.99
10
0740A
:TRANsition
[:LEADing]
0 to 99.99
10
0741A
:TRAiling
0 to 99.99
10
0742A
:SINC
:NCYCle
4 to 100
10
0790A
:GAUSsian
:EXPonent
1 to 200
10
0750A
:EXPonential
:EXPonent
-100 to 100
-10
0760A
:DC
[:VOLTage]
-100 to 100
100
0770A
:AMPLitude
-8 to 8
5
0046A
FUNCtion:SHAPe {SINusoid|TRIangle|SQUare|PULSe|RAMP|SINC|
GAUSsian|EXPonential|DC|NOISe}(?)