User manual

Table Of Contents
3152B User Manual Publication No. 980935 Rev. M
5-42 Programming Reference Astronics Test Systems
Parameters
Name
Type
Default
Description
EXTernal
Discrete
EXT
Selects the front panel TRIG IN input as the trigger
source.
INTernal
Discrete
Activates the built
-in inter
nal trigger generator. BUS and
external trigger
s
are ignored. The period of the internal
trigger is programmable and can be used to replace an
external trigger source.
TTLTrg<0 to 7>
Discrete
Selects a backplane trigger line
as the source for the
trigger
input. To avoid hardware conflicts, make sure
that no more than one instrument is programmed to
output trigger signals on any specific TTLT line.
ECLTrg1
Discrete
Selects the backplane ECL trigger line number 1 as the
source for the trigger input.
BUS
Discrete
Selects the remote controller as the trigger source. Only
software commands are accepted;
b
ackplane and front
panel signals are ignored
.
Response
The 3152B returns EXT, INT, TTLT<n>, or BUS depending on the selected trigger advance source setting.
TRIGger:SLOPe {NEGative|POSitive}(?)
HS Syntax: 1160A<
0|1>(?)
Description
The trigger slope command selects the
polarity of the trigger signal as the trigger source. This affects both
Transition (edge) mode and Level mode.
Transition mode: If the
POS option is selected, then positive-going transitions toggle the gate open or closed.
If the NEG option is selected, then negative
-going transitions toggle the gate open or closed. Two transitions
in the same direction are required to gate the output on
and then off. The trigger slope command affects the
generator when in an interrupted run mode, which can be selected using the INIT:CONT OFF command.
Level mode: If the POS option is selected, then a trigger signal above the threshold opens the gate, and
a
trigger signal below the threshold closes the gate. If the NEG option has been selected, then a trigger signal
below the threshold opens the gate, and a trigger signal above the threshold closes the gate.
Parameters
Name
Type
Default
Description
NEGat
ive
Discrete
Transition (edge) mode: Selects the negative going
edge
as the trigger source.
Level mode: A
negative level (below the trigger level
setting) opens the gate.
POSitive
Discrete
POS
Transition (edge) mode:
Selects the positive going edge
as th
e trigger source.
Level mode: A
positive level (above the trigger level
setting) opens the gate.
Response
The 3152B returns POS or NEG depending on the selected trigger slope setting.