3152B User Manual Publication No. 980935 Rev. M
5-40 Programming Reference Astronics Test Systems
<time>
0, 10 to
2e6
Numeric
(integer only)
0
“0” turns off the delayed trigger function. Delay is
programmed in sample clock increments, so expect the
delay time to change if you modify your sample clock
setting.
Response
The 3152B returns the current trigger delay time value in sample clock cycles.
TRIGger:DELay {OFF|ON|0|1}(?)
HS Syntax: 0188A<0|1>(?)
Description
Use this command to
toggle the delayed trigger feature. The trigger delay command affects
the generator
only after it has been programmed to operate in
interrupted run mode. Set the 3152B to interrupted run mode
using the
INIT:CONT OFF command.
Range
Type
Default
Description
0
-1
Discrete
0
Toggles
delayed trigger mode.
Response
The 3152B returns 0 or 1 depending on the selected option.
TRIGger:DELay:Time <time>(?)
HS Syntax: 0156A<float>(?)
Description
The trigger delay time param
eter defines the time that elapses between the receipt of a valid trigger signal
until
the initiation of the first output waveform. Trigger delay can be turned off and on using the TRIG:DEL
<0|1> command. The
trigger delay time command affects the generator only after it has been programmed to
operate in interrupted run mode. Modify the 3152B to interrupted run mode using the INIT:CONT OFF
command.
Parameters
Name
Rang
e
Type
Default
Description
<time>
100e
-9 to 20
Numeric
100e
-9
Programs the trigger delay time. Programming
resolution is 20 ns across the range.
Response
The 3152B returns the current trigger delay time value.
TRIGger:GATE:MODE {LEVel|TRANsition}(?)
HS Syntax: 0157A<0|1>(?)
Description
Sets
the 3152B gate mode to TRANsition (start and stop pulses) or LEVel (open gate on high or low level)
.
The trig:slop command defines the polarity of both
modes. Select the source as either the front panel TRIG IN
connector or one of the backplane trigger lines.
Note that you can control the trigger level threshold,
but only if
you can only do it if you
are using the front panel input. This command affects the 3152B only after it is
set to
non
-continuous mode (INIT:CONT OFF).