User manual

Table Of Contents
Publication No. 980935 Rev. M 3152B User Manual
Astronics Test Systems Programming Reference 5-39
Response
The 3152B will respond to a remote trig command if INIT:CONT 0 and TRIG:SOUR BUS are both true.
TRIGger:BURSt {OFF|ON|0|1}(?)
HS Syntax: 1120A<0|1>(?)
Description
Th
is command toggles counted burst run mode on and off. This command affects the 3152B
when it is in
INIT:CONT 0 mode.
Parameters
Range
Type
Default
Description
0
-1
Discrete
0
“1”
enables counted burst run mode. “0” turns
burst run
mode off. Burst count is programmable using the
TRIG:BURS:COUN command.
Response
The 3152B returns 0 or 1 depending on the selected option.
TRIGger:BURSt:COUNt <burst>(?)
HS Syntax: 1121A<long>(?)
Description
This function sets the number of
burst cycles to be output when Burst Mode is enabled. Use the INIT:CONT
OFF;:TRIG:BURS ON
commands to select Burst Mode.
Parameters
Name
Ran
ge
Type
Default
Description
<burst>
1 to 1M
Numeric
(integer only)
1
Programs the burst count.
Response
The 3152B returns the current burst count value.
TRIGger:DELay <sample clock cycles>(?)
HS Syntax: 1131A<long>(?)
Description
The trigger delay
time parameter defines the time that elapses between the receipt of a valid trigger signal
until
the initiation of the first output waveform. Trigger delay is turned off using the TRIG:DEL 0 command.
The trigger delay time command will affect the generato
r only after it has been programmed to operate in
interrupted run mode.
Set the 3152B into interrupted run mode using the INIT:CONT OFF
command. Note
that this command is left in the commands list for compatibility with the model 3152A. The delay time is
p
rogrammed in sample clock increments. For new applications the TRIG:DEL:TIME
command is
recommended because it allows
you to program delays in time units instead of in sample clock periods.
Parameters
Name
Range
Type
Default
Description