User manual

Table Of Contents
3152B User Manual Publication No. 980935 Rev. M
5-38 Programming Reference Astronics Test Systems
Table 5-4, Run Mode Commands
Keyword
Parameter Form
Default
3152A
HS
:INITiate
[:IMMediately]
:CONTinuous
OFF | ON | 0 | 1
1
1110A
:TRIGger
[:IMMediate]
11000
:BURSt
[:STATe]
OFF | ON | 0 | 1
0
1120A
:COUNt
1 to 1000000
1
1121A
:DELay
0, 10 to 2e6 (points, even numbers)
0 (0 = OFF)
1131A
[:STATe]
OFF | ON | 0 | 1
0
0188A
:TIMe
100e-9 to 20
100
0156A
:GATE
:MODe
LEVel | TRANsition
LEV
0157A
[:STATe]
OFF | ON | 0 | 1
0
1150A
:LEVel
-10 to 10
1.6
1180A
:SOURce
[:ADVance]
EXTernal | INTernal | TTLTrg<n> | BUS
EXT
1140A
:SLOPe
NEGative | POSitive
POS
1160A
:RETRigger
[:STATe]
OFF | ON | 0 | 1
0
0187A
:TIMe
100e-9 to 20
100
0161A
:TIMer
1e-6 to 20
15e-6
1170A
INITiate:CONTinuous {1|0|ON|OFF}(?)
HS Syntax: 1110A<1|0>(?)
Description
Toggles
the 3152B to operate in either continuous mode (INIT:CONT ON | 1) or in interrupted run mode
(INIT:CONT OFF | 0)
. Trigger subsystem commands affect the 3152B only after it is set for
interrupted run
mode.
Parameters
Range
Type
Default
Description
1
-0
Discrete
1
“1”
selects continuous run mode. “0” selects interru
pted
run mode.
In INIT:CONT 0 mode
, you can program the
3152B to operate in triggered, gated, or counted burst
run modes.
Response
The 3152B returns 1 or 0 depending on the selected option.
TRIGger
HS Syntax: 11000
Description
Use this command to
send a software trigger to the 3152B. The *trg command has
the same effect. This
command will affect the 3152B
while it is running in interrupted run mode (INIT:CONT 0) and only when the
selected trigger source is BUS.