3152B User Manual Publication No. 980935 Rev. M
5-34 Programming Reference Astronics Test Systems
Description
This command queries the frequen
cy at the trigger input. The same trigger is also used for phase locking to an
ext
ernal signal. This command returns the frequency value of the external signal only when the instrument is in
PLL mode and locked.
Response
The 3152B
measures and returns the current frequency applied to the trig/pll input
. If no signal is applied to
the trigger input, the response will be 0. The returned value will be in
scientific notation
(for example: 100mHz
would be returned as 100e-3.
FREQuency {<freq>|MINimum|MAXimum}(?)
HS Syntax: 0400A<float>(?)
Description
M
odifies the frequency of the standard waveform and is specified in units of H
ertz (Hz). It has no affect on
arbitrary waveforms.
Parameters
Name
Range
Type
Default
Description
<freq>
10
0e-6 to
100e6
Numeri
c
1e6
Sets
the frequency of the standard waveform in units of
Hz. The frequency command can be used with
resolutions up to 14 digits.
<MINimum>
Discrete
Sets
the frequency of the standard waveform at the
lowest possible
frequency (100e-6).
<MAXimum>
Discrete
Sets
the frequency of the standard waveform to the
highest possible frequency (100e6).
Response
The
3152B returns the current frequency value. The returned value will be in scientific notation
(for example:
100mHz would be returned as 100e-3 positive numbers are unsigned).
FREQuency:RASTer {<sclk>|MINimum|MAXimum}(?)
HS Syntax: 0410A<float>(?)
Description
M
odifies the sample clock frequency of the arbitrary waveform in units of samples per second. It has no affect
on standard waveforms.
Parameters
Name
Range
Type
Default
Description
<sclk>
10
0e-3 to
25
0e6
Numeric
1e6
Sets
the sample clock frequency of the arbitrary and
sequenced waveform
in units of samples per second.
The sample clock can be programmed with
a resolution
of
up to 14 digits.
<MINimum>
Discrete
Sets
the sample clock frequency to the lowest possible
frequency (100e
-3).