User manual

Table Of Contents
3152B User Manual Publication No. 980935 Rev. M
5-32 Programming Reference Astronics Test Systems
Name
Range
Type
Default
Description
<width>
4 to
500
(2 to
500 in
legacy mode)
Integer
4
(2 in
legacy
mode)
For s
tandard waveform: Sets the SYNC duty cycle. The
percentage duty cycle is <width>/20. For example, if
<width>
= 500, then the duty cycle is 500/20, or 25%.
The <width> values are in increments of 4 (2 in legacy
mode).
For a
rbitrary (user) waveform: Sets the SYNC pulse
width
. The <width> value specifies the time expressed
in
waveform points, in increments of 4 (2 in legacy
mod
e).
Response
The 3152B returns the current SYNC width value.
OUTPut:TRIGger:SOURce {BIT|LCOM|SSYN|HCL|PULS|ZERO|INTernal|
EXTernal}(?)
Description
The TTLTRG signals, when enabled and placed on the backplane, can be asserted with signals coming from
a number of sources. Use this command to assign the source for the active TTLTRG line.
HS Syntax: 0220A<1|2|3|4>(?)
Parameters
Name
Type
Default
Description
BIT
Discrete
BIT
Generates a trigger signal at a designated point on the
waveform. The trigger position is programmed using the
outp:sync:pos command. The same command sets the
position of the trigger output and the position of the
SYNC output.
LCOMplete
Discrete
Generates a single trigger signal in sequence
d
mode
synchronous with
the active segment in the sequence.
SSYNc
Discrete
S
imilar to the bit option except that
it in triggered or
gated modes
it eliminates the ±1 clock jitter between th
e
trigger and the output signal.
HCLock
Discrete
G
enerates a sync output pulse that has a
50% duty
cycle
and the same period as the output waveform.
PULSe
Discrete
Similar to the bit option except that
the wid
th of the sync
pulse
in 4 point increments from 4 to n-
8 points (2 point
increments from 2 to n
-4 in legacy mode)
. The position
of the pulse is programmed using the outp:sync:pos
command and
the
width is programmed using the
outp:sync:wid command.
ZEROcross
Discrete
Zero crossing is a special mode where the sync signal
remains low as long as the output waveform level is
negative (below 0 V)
and changes
to high when the
output level becomes positive
(above 0 V).
INTernal
Discrete
This type selects the internal generator as the source.