Publication No. 980935 Rev. M 3152B User Manual
Astronics Test Systems Programming Reference 5-29
The 3152B returns NONE, 25M, 50M, 60M, or 120M depending on the filter currently connected in series with
the output signal.
OUTPut:FILTer {OFF|ON|0|1}(?)
HS Syntax: 0202A<0|1>(?)
Description
This command toggle
s selection of the filter selected by
the OUTP:FILT command. Note that the filters are
not accessible during standard sine waveform output.
In order to control
filters, change the output waveform
function to another waveform type to allow this command to execute.
Parameters
Range
Type
Default
Description
0
-1
Discrete
0
Toggles selection of the selected
output filter
Response
The 3152B returns 1 if a filter is enabled or 0 if a filter is disabled
OUTPut:LOAD <load>?
HS Syntax: 0012A<integer>(?)
Description
This command specifies the load impedance to be applied to the 3152B main output.
Parameters
Name
Type
Default
Description
<load>
Numeric (integer
only)
50
Specifies
the load impedance that will be applied to the
3152B outputs in units of
. The default setting is 50
.
The range of load impedance is 50
to 1 M
. Accurate
specification
of the load impedance is
required to
automatically apply the programmed
amplitude level
at
the load.
OUTPut:LOAD
:CAL {50|1M}(?)
HS Syntax: 0013A<0|1>(?)
Description
This
command specifies the table of calibration constants to activate. This provides maximum accuracy for
either 50 or high-impedance (>1 MΩ) loads.
Parameters
Name
Type
Default
Description
50
Discrete
50
Specifies
that the 50
calibration table will be applied
to output offse
t and amplitude settings.
1M
Discrete
Specifies
that the 1 MΩ
calibration table will be applied
to output offset and amplitude settings.
Response
The 3152B returns 50 for 50 Ω or 1M for 1 MΩ.