Publication No. 980935 Rev. M 3152B User Manual
Astronics Test Systems Programming Reference 5-27
Response
The 3152B will return ADJ, ECLT or LBUS depending on the current instrument couple path assignment.
INSTrument:COUPle:SLAVe:DELete <LAN_address>
HS Syntax: 00040<string>(?)
Description
This command will delete a designated slave instrument from
a synchronized multi-instrument
system list.
This command is associated with LAN operation only.
Parameters
Name
Type
Default
Description
<LAN_address>
String
Will
remove a designated instrument (specified by IP
address)
from a synchronized multi-instrument system
list. Co
ntact your IT department, if you are not sure how
to connect to a
LAN address.
INSTrument:COUPle:SLAVe:INSert <3152B>,<LAN_address>
HS Syntax: 00051<3152B,string>(?)
Description
This command will add a designated slave instrument to a synchronized
multi-instrument
system list. This
command is associated with LAN operation only.
Parameters
Name
Type
Default
Description
<3152B>, <LAN_address>
String
Specifying the correct model number is crucial for
correct assignme
nt of the instrument designator and
for
selection of
the correct instrument number for the
INST:SEL command.
INSTrument:COUPle:STATe {OFF|ON|0|1}(?)
HS Syntax: 0006A<0|1>(?)
Description
This command will turn the 3152B coupled state on and off.
Parameters
Range
Type
Default
Descripti
on
0
-1
Discrete
0
Turn
s coupled
mode on and off. Note that this command
must be applied to the master instrument only
,
otherwise,
slave instruments will not couple. To select the master
instrument use the INST:SEL 1 command.
Response
The 3152B returns 1 if the coupled state is on or 0 if the couple state is off.