User manual

Table Of Contents
3152B User Manual Publication No. 980935 Rev. M
5-8 Programming Reference Astronics Test Systems
3152B Legacy
Commands
The 3152B is a modern and updated version of the Model 3152A
employing the latest technology and component improvements
.
A
lthough 100% backwards compatibility was the basi
s for the new
3152B design, some of the 3152A functionality could not be
duplicated exactly. Table 5
-
1 lists all of the 3152B SCPI commands
and highlights the areas where it differs from the original 3152A
design. Expect that commands that are not compatible with the old
3152A will generate errors
when the instrument is in legacy mode.
Table 5-1, 3152B SCPI Command Summary for 3152A Emulation
Keyword Parameter Form Default 3152A HS
Instrument Control Commands
:INSTrument
:MODE NORMal | FAST NORM
0150A
Run Mode Commands
:INITiate
[:IMMediately]
:CONTinuous OFF | ON | 0 | 1 1
1110A
:TRIGger
[:IMMediate]
11000
:BURSt
[:STATe] OFF | ON | 0 | 1 0
1120A
:COUNt 1 to 1000000 1
1121A
:DELay 0, 10 to 2e6 (even numbers) 0 (0 = OFF)
1131A
:GATE
[:STATe] OFF | ON | 0 | 1 0
1150A
:LEVel -10 to 10 0
1180A
:SOURce
[:ADVance] EXTernal | INTernal | TTLTrg<n> | ECLTrg1 EXT
1140A
:SLOPe NEGative | POSitive POS
1160A
:TIMer 1e-6 to 20 15e-6
1170A