User manual

Table Of Contents
3152B User Manual Publication No. 980935 Rev. M
4-70 ArbConnection Astronics Test Systems
Pulse Example,
Section
4
The third pulse section is now
complete. We are ready now to start
building the forth section of the pulse as shown in Figure 4-54. Use
the Edit menu to select the Append Section operation.
A new
section number will appear but its fields will be initially empty to the
right of the section identifier.
Before you start entering values in
to this section, note that there is
only one linear transition required for this section that will start from
the last point of the previous section and will connect to the start
point of the next section. Therefore, select the Time/Level Points
option in the Pulse Train Design Format. You are now ready to start
programming values. Type the section entries as shown
in the
figure.
Figure 4-54, Building Section 4 of the Pulse Example
Pulse Example,
Section
5
The fourth pulse section is complete so we are now
ready to start
building the fifth and final section of the pulse as shown in Figure 4-
55. Use the Edit menu to select the Append Section operation. A
new section number will appear but its fields will be initially empty to