3152B User Manual Publication No. 980935 Rev. M
4-36 ArbConnection Astronics Test Systems
demote a master to a servant. Note that this operation affects a
line that is currently highlighted.
Path (LBUS/ECLT) is used for selecting the connection path.
LBUS specifies a VXI local bus connection and requires that the
master is plugged into the leftmost position and all servants are
plugged into adjacent slots to the right to the master module. ECLT
specifies the VXI backplane ECLTrg line synchronization mode.
For this mode, the location of the master and servant units in the
chassis is not crucial because the trigger lines run across the
backplane through all slots.
Apply used to prepare the instruments for the synchronization
sequence. This button must precede the activation of the
synchronization process.
Activate Once this button is pressed,
the instruments will
synchronize in the selected groups and the master will control the
timing of the servant modules.
Close terminates the current session but does not change the
synchronization status.
The System Panels
The System tab (Figure 4-26) provides access to a group of panels
that control some general system parameters and provides access
to calibration
. There are two panels in this group: General/System,
which provides access to some system commands, utilities and
filters; and Calibration, which provides access to the remote
calibration utility. Note however, that access to the calibration panel
is permitted to qualified service personnel and requires a user name
and password. Information on how to access the calibration panel is
given in Chapter 7.
Figure 4-26, System Panels
General/Filters
The General/Filters panel (Figure 4-27) provides access to some
general system common commands, allows read back of