Astronics Test Systems ArbConnection 4-33
Figure 4-22, X-Instrument Synchronization Pool List
Group is an edit field which is used for grouping one or more
instruments into a set of instruments that share synchronization
properties.
State identifies the master or servant property of an instrument.
Note that the first instrument in the group list is always set
automatically as the master. If you want it as servant, you can use
the Move Up and Move Down buttons to move the module higher
or lower in the synchronization hierarchy.
Model
shows all instruments from the 3152B family that
ArbConnection detected in the chassis. Leave the group field blank
if you do not wish to synchronize a particular module.
Address shows the logical address associated with the listed
module. The address is set using a DIP switch which can be
accessed when the module is removed from the chassis.
Chan (Channel) this field is relevant for the 3100-3152B-3152B
model which has two 3152B units installed in a single VXI slot. In
this case, each instrument operates as a stand-alone generator but
also can be configured as a dual-channel instrument. The example
in Figure 4-22 shows this instrument with address 13 installed in
slot 3. Channel 1 is always the top instrument and channel 2 is
installed below and if selected as a group, by default, channel 1
becomes the master module. There is no way to exchange channel
designation but if you move channel 2 up, its state is re-configured
to master and channel 1 to servant.
Slot
shows the slot number where ArbConnection found a
specific module. The location of the module is extremely important
because it defines how it may be synchronized, as explained in the
Path description below.
Pathdefines the connection between synchronized modules.
This field has three options: ADJ, LBUS and ECLT.