4-28 ArbConnection Astronics Test Systems
modulation carrier waveform.
Baseline
The Baseline parameter affects the output
characteristics in one of the interrupted run modes (i.e., triggered,
burst). In this case this parameter defines where the signal idles
between triggers. There are two options: CW and DC
. The DC
option will set the idle state to a DC level, meaning that in between
triggers, the output resides on a DC
level and generates modulation
when a trigger is accepted. The CW is similar except the signal
idles on the pre-trigger CW frequency setting
, executes the
modulation upon receipt of a legal trigger signal and returns to
continuous CW frequency output.
Amplitude Hop
The Amplitude Hop group contains parameters that control the
amplitude hop function. These are: hop data, dwell control, dwell
time and marker position.
Hop Data
The Hop Data button in the Ampl Hop group provides
access to the data string that controls the sequence of amplitude
hops. The hop data table contains a list of amplitude levels that the
output steps through the amplitude levels of
as programmed in the
hop data table.
Fixed Hold The hold parameter determines how long each step
of amplitude dwells
on this setting before it will step to the next
amplitude setting. By selecting
Fixed Hold, the hold time remains
constant throughout the entire hop table.
Figure 4-18, Amp/Freq Hop Panel
Variable Hold The Variable Hold parameter determines how long
each step of amplitude dwells before stepping to the next amplitude
setting. By selecting Variable Hold
, the hold time changes
automatically from one step to the next
depending on the hold time
value that is affixed to the hop step. The values can be
programmed in the HOP Data table.