User manual

Table Of Contents
3152B User Manual Publication No. 980935 Rev. M
4-26 ArbConnection Astronics Test Systems
Figure 4-17, FSK/PSK/ASK Modulation Panel
“0/1” Frequency
In FSK, the carrier waveform (CW) has two
frequencies: an initial fre
quency level which is set by the “0”
f
requency parameter and shifted frequency which is set by the “1”
frequency. The control data table has a list of “0” and “1” values that
flag when the frequency shifts from base to shifted frequency.
Baud The baud parameter sets the rate at
which the generator
steps through the sequence of the FSK Control Data bits.
Marker Index The marker index programs a step
in the control
data string to output a pulse at the SYNC output connector. The
SYNC State button must be t
urned on to generate the FSK marker
output.
PSK
The PSK group contains parameters that control the phase shift
keying function. These are
: control data, “0” and “1” frequencies,
baud rate, and marker position.
Control Data The Control Data button in the
PSK group provides
access to the data string that controls the sequence of base phase
and shifted phase. It contains the list of “0” and “1” values that the
output will repeatedly follow for its phase
shift keying sequence
advance as programmed.
“0/1” Phase
In PSK, the carrier waveform (CW) has two phase
settings: an initial phase which is set by the “0” Phase parameter
and shifted phase which is set by the “1” Phase. The control data
table has a list of “0” and “1” values that flag when the phase shifts
from base to shifted phase.
Baud The baud parameter sets the rate at
which the generator
steps through the sequence of the PSK Control Data bits.
Marker Index The marker index programs a step
in the control
data string to output a pulse at the SYNC o
utput connector. The
SYNC State button must be turned on to generate the PSK marker