Specifications
Maintenance and Performance Checks 5-13
NOTE
If not otherwise specified, configure the instrument to factory
defaults before each adjustment by sending the *RST
command. Always connect the output BNC connector through
a 50SS feedthrough termination.
Pulse Response
Adjustment
Equipment: Oscilloscope (2465B), 20dB Attenuator
1. Configure the Model 3152 as follows:
Function: Square
Frequency: 1 MHz
Amplitude: 10 V
Remote Commands::RES
:OUTP ON
:FUNC:SHAP SQU
:FREQ 1e6
:VOLT 10
2. Connect the Model 3152 output through 20dB attenuator
to the oscilloscope input.
3. Set the oscilloscope input impedance to 50S.
4. Set the oscilloscope vertical gain and time base, and
adjust R36 and C19 for the best pulse response.
5. Change the Model 3152 frequency setting to 10.00 KHz.
6. Set the oscilloscope vertical gain and time base, and
adjust R22 for best pulse flatness.
7. Change the Model 3152 frequency setting to 1.000 KHz
and retouch R22 for best flatness in this range. Repeat
Steps 5 and 6 until the best result is obtained in both
steps.