Specifications
Maintenance and Performance Checks 5-10
Burst
1. Configure the Model 3152 as follows:
Function: Sine wave
Frequency: 1.000 KHz
Operating Mode: Burst
Remote Commands: :RES
:OUTP ON
:FREQ 10e3
:INIT:CONT OFF
:BURST:STAT ON
:TRIG:COUN 10
2. Set the oscilloscope and verify that the Model
3152 outputs a burst of 10 complete output
waveforms. Remove the external pulse generator
connection from the Model 3152 for the next test.
PLL Characteristics Accuracy: 5.4 x frequency (in MHz) + resolution / 2 ±5E
Equipment: Oscilloscope, Counter, Pulse/Function Generator,
Feedthrough Terminator, 3 BNC Cables, 50S “T” Connector
1. Configure the Model 3152 as follows:
Function: Square wave
Mode: Front Panel PLL
Frequency: 1.000 KHz
Amplitude: 2 V
Trigger Level: 0 V
Remote Commands: :RES
:OUTP ON
:FUNC:SHAP SQU
:PHAS2:LOCK ON
:TRIG:LEV 0
:VOLT 2
2. Make the following connections:
a. Connect function generator output to the input of the
50S “T” Connector.
b. Connect one end of the 50S “T” Connector to the
3152 TRIG IN and the other end to the counter
Channel A input.
c. Connect 3152 OUT to the counter Channel B input.
3. Set counter to phase measurements. Terminate counter
inputs with 50S.