Specifications
Maintenance and Performance Checks 5-6
Equipment: DMM, 50S feedthrough termination
1. Configure the Model 3152 as follows:
Function: Sine
Frequency: 1.000 MHz
Amplitude: 1.61 V
Remote Commands::RES
:OUTP ON
:FUNC SIN
:FREQ 1e6
:VOLT 1.61
:VOLT:OFFS 5
2. Set the DMM to DCV measurements
3. Connect the Model 3152 output to the DMM input.
Terminate the output with a 50S feedthrough termination.
Set the Model 3152 offset and verify the DMM reading as
follows:
OFFS SETTING DMM READING
±5.000 V ±5.070 V to ±4.930 V
4. Change the Model 3152 AMPL setting to 161 mV. Set
VOLT:OFFS and verify the DMM reading as follows:
OFFS SETTING DMM READING
±500.0 mV ±507.0 mV to ±493.0 mV
5. Change the Model 3152 AMPL setting to 16.1 mV. Set
VOLT:OFFS and verify the DMM reading as follows:
OFFS SETTING DMM READING
±50.00 mV ±50.70 mV to ±49.30 mV
Squarewave
Characteristics
Specified Transition Time: 5 ns typical (10% to 90% of amplitude)
Specified Aberration: <5% of amplitude
Equipment: Oscilloscope, 20dB attenuator
1. Configure the Model 3152 as follows:
Function: Square wave
Frequency: 1.000 MHz
Amplitude: 10 V
Remote Commands::RES
:OUTP ON
:FUNC:SHAP SQU
:FREQ 10e6
:VOLT 10