Specifications

Maintenance and Performance Checks 5-5
3. Connect the Model 3152 output to the counter’s input.
Change the frequency setting as required for the test and
verify the reading on the counter display as follows:
3152 SETTING COUNTER READING
10.00000 Hz9.999000 Hz - 10.00100 Hz
100.0000 Hz99.99000 Hz - 100.0100 Hz
1.000000 KHz 999.9000 Hz - 1.000100 KHz
10.00000 KHz 9.999000 KHz - 10.00100 KHz
100.0000 KHz 99.99000 KHz - 100.0100 KHz
1.000000 KHz 999.9000 KHz - 1.000100 KHz
10.00000 MHz 9.999000 MHz - 10.00100 MHz
50.00000 MHz 49.99500 MHz - 50.00500 MHz
Amplitude Accuracy Accuracy Specifications (1 KHz):
±(1% of reading +200 µV), to 160 mV
±(1% of reading +2 mV), to 1.6 V
±(1% of reading +20 mV), to 16 V
Equipment: DMM, 50S feedthrough termination
1. Configure the Model 3152 as follows:
Function: Sine
Frequency: 1.000 KHz
Amplitude: 15 V
Remote Commands::RES
:OUTP ON
:FUNC:SHAP SIN
:FREQ 1000
:VOLT 15
2. Set the DMM to ACV measurements (RMS).
3. Connect the Model 3152 output to DMM input. Terminate
the output with a 50S feedthrough termination. Set the
amplitude and verify DMM reading as follows:
3152 SETTING DMM READING
15.00 V 5.373 V - 5.230 V
1.500 V 537.3 mV - 523.0 mV
150.0 mV 53.73 mV - 52.30 mV
DC Offset
Characteristics
Accuracy Specifications:
±(1% of reading +1% of amplitude +200 µV), ±80 mV window
±(1% of reading +1% of amplitude +2 mV), ±800 mV window
±(1% of reading +1% of amplitude +20 mV) ±8 V window