Specifications
Maintenance and Performance Checks 5-4
Recommended Test
Equipment
Recommended test equipment for troubleshooting, calibration and
performance checking is listed below. Test instruments other than
those listed may be used only if their specifications equal or exceed
the required characteristics.
Test Instrument & Accessories Minimum Specification
Universal Counter 100 MHz, 10 ppm accuracy
DMM ACV, 0.2%; DCV, 0.1% accuracy
Distortion Analyzer 10 Hz to 100 KHz
Spectrum Analyzer 1 GHz bandwidth, 1 KHz resolution
Pulse/Function Generator 20 MHz, 10 ns pulsewidth
Oscilloscope 2 channels, 400 MHz analog bandwidth
Feedthrough termination 50S, 1%
BNC Cables 1 meter long, 50S impedance
20dB feedthrough attenuator 50S, 1%
âTâ BNC Connector 50S, 1%
Performance
Check Procedures
Use the following procedures to check the Model 3152 against the
specifications. A complete set of specifications is listed in Appendix
A. The following paragraphs show how to set up the instrument for
the test, what the specifications for the tested function are, and what
acceptable limits for the test are. If the instrument fails to perform
within the specified limits, the instrument has to be calibrated or
tested to find the source of the problem.
Frequency Accuracy
Accuracy Specifications: 0.01% of reading to 50.00 MHz
Equipment: Counter
1. Configure the Model 3152 as follows:
Function: Square
Frequency: 10 Hz
Amplitude: 1 V
Remote Commands::RES
:OUTP ON
:FUNC:SHAP SQU
:FREQ 10
:VOLT 1
2. Set the counter to frequency measurement.